Pre Column Installation Check List

Proper column installation is essential in obtaining the highest performance and lifetime from a capillary column.

- Cool down the injector and detector

- Inspect gas filters and replace saturated traps                            "Click to GC - Gas filters - Filtres GC" >>>

- Check the injector and the detector, Clean or replace all dirty or corroded parts

- Change the injection Septum, the inlet liner and the injector seals

                                                                                  "Click to GC Injection Septa" >>>

<<< "Click to Injection Liners"

- Place the nut and appropriate ferrule* on the column and carefully cut the column end (up to 10cm) using a ceramic wafer or a diamond tipped pencil or a sapphire cutting tool.

Column Inner diameter Ferrule Inner diameter    <<< "Click to GC- Capillary ferrules"
0.10-0.18-0.20-0.25 mm 0.4 mm
0.32 mm 0.5 mm
0.45-0.53 mm 0.8 mm
   

The optimal insertion distance of the column into the injector is different for each model of GC (mentioned in your instrument manual)

- Turn on the carrier gas to the flow rate or inlet pressure recommended for the column. Confirm the presence of column flow by immersing the column outlet in a vial of solvent and flush the column at ambient temperature at least 5 minutes for 25-30m column and 10 minutes for 50-60m column.

Approximate Head Pressure (PSI)
  0.18mm 0.20mm 0.25mm 0.32mm 0.45mm 0.53mm
10m 5-10          
12m   10-15        
15m     8-12 5-10   1-2
20m 10-20          
25m   20-30        
30m     15-25 10-20 3-5 2-4
40m 20-40          
50m   40-60        
60m     30-45 20-30 6-10 4-8
75m         8-14 5-10
105m           7-15

 - Install the column into the detector as described in the instrument manual.
Set the detector gases and temperatures to proper settings

 - Inspect for leaks all connections using a leak detector

                                                            
"Click to GC-Gas filters - leak detector" >>>

- Verify the carrier gas flow, set the split vent, septum purge and any other application gas rates as appropriate

  <<< "Click to GC-Gas filters - Gas flowmeter"

- Confirm carrier gas flow and proper column installation

- Condition the column 20°C above the final analysis temperature of your method. Do not exceed the column’s maximum operating temperature

- To check for instrument performance, analyse a column test mix for a new method or a known standard to confirm proper column and system performance.

- Accurately set the carrier gas velocity (recommended average linear velocity 30 to 40 cm/sec for Helium, 50 to 80 cm/sec for Hydrogen)
Set the oven to the initial temperature of the temperature program or to the isothermal value. Inject 1-2µl of the appropriate non-retained compound. Adjust the head pressure until the desired average liner velocity is obtained.

- Bleed test : acquire a blank run (no injection) using a temperature program. Starting at 40–50°C, ramp 10-20°/min, then hold for 10-15 minutes at the conditioning temperature.

- Run test mix : Inject a test mixture to measure system performances . The test mixture is useful as a comparative and diagnostic sample.